Development and validation of an ex vivo porcine model of functional tricuspid regurgitation
Background and Aim: Ex vivo models of functional tricuspid regurgitation (FTR) are needed for pre-clinical testing of novel surgical and interventional repair strategies, but current options are costly or have not been formally validated. The objective of this research was to create and validate an ex vivo model to test novel repair methods for FTR.
Methods: In explanted porcine hearts, the right atrium was excised to visualize the tricuspid valve. The pulmonary artery and aorta were clamped and cannulated, the coronary arteries ligated, and the right and left ventricles statically pressurized with air to 30 mmHg and 120 mmHg, respectively. FTR was induced by increasing right ventricular pressure to 80 mmHg for 3 h, which resulted in progressive tricuspid annular enlargement, right ventricular dilation, papillary muscle displacement, and central tricuspid malcoaptation. A structured light scanner was used to image the 3D topography of the tricuspid valve in both the native and FTR state, and images were exported into scan-to-computer-aided design software, which allowed for high-resolution 3D computational reconstruction. Relevant geometric measurements were sampled including annular circumference and area, major and minor axis diameter, and tenting height, angle, and area. Geometric measurements from the ex vivo model were compared to clinical transthoracic echocardiographic (TTE) measurements using two-sample t-tests.
Results: A total of 12 porcine hearts were included in the study. Annular measurements of the native valve were comparable to published TTE data, except for the minor axis diameter, which was shorter in the ex vivo model (2.5 vs. 3.1 cm, P = 0.007). Induction of FTR in the ex vivo model resulted in annular enlargement (FTR vs. native: circumference 13.7 vs.11.8 cm, P = 0.012; area 14 vs. 11 cm2, P = 0.011). Ex vivo leaflet measurements in both the native and FTR model differed from published TTE data, but demonstrated comparable directional changes between the native and regurgitant states, including increased tenting height, area, and volume.
Conclusion: The ex vivo pneumatically-pressurized porcine model closely recapitulates the geometry of both the native and regurgitant tricuspid valve complex in humans and holds promise for testing novel FTR repair strategies.
Relevance for Patients: Currently available interventions for the tricuspid valve have a risk of permanent conduction abnormalities and are insufficient in addressing tricuspid disease for a subset of patients. This ex vivo model provides a platform for testing of novel interventions that address the deficiencies of current tricuspid therapies.
[1] Topilsky Y, Khanna A, Le Tourneau T, Park S, Michelena H, Suri R, et al. Clinical Context and Mechanism of Functional Tricuspid Regurgitation in Patients with and without Pulmonary Hypertension. Circ Cardiovasc Imaging 2012;5:314-23. doi: 10.1161/CIRCIMAGING.111.967919
[2] Gammie JS, Chu MW, Falk V, Overbey JR, MoskowitzAJ, Gillinov M, et al. Concomitant Tricuspid Repair in Patients with Degenerative Mitral Regurgitation. N Engl J Med 2022;386:327-39. doi: 10.1056/NEJMoa2115961
[3] Adkins A, Aleman J, Boies L, Sako E, Bhattacharya S. Force Required to Cinch the Tricuspid Annulus: An Ex-vivo Study. J Heart Valve Dis 2015;24:644-52.
[4] Stock S, Bohm H, Scharfschwerdt M, Richardt D, Meyer-Saraei R, Tsvelodub S, et al. Ex Vivo Hydrodynamics after Central and Paracommissural Edge-to-Edge Technique: A Further Step Toward Transcatheter Tricuspid Repair? J Thorac Cardiovasc Surg 2018;155:949-55. doi: 10.1016/j.jtcvs.2017.10.068
[5] Amini Khoiy K, Biswas D, Decker TN, Asgarian KT, Loth F, Amini R. Surface Strains of Porcine Tricuspid Valve Septal Leaflets Measured in Ex Vivo Beating Hearts. J Biomech Eng 2016;138:111006. doi: 10.1115/1.4034621
[6] Amini Khoiy K, Asgarian KT, Loth F, Amini R. Dilation of Tricuspid Valve Annulus Immediately after Rupture of Chordae Tendineae in Ex-Vivo Porcine Hearts. PLoS One 2018;13:e0206744. doi: 10.1371/journal.pone.0206744
[7] Nguyen YN, Tay EL, Kabinejadian F, Ong CW, Ismail M, Leo HL. Ventricular Vortex Loss Analysis Due to Various Tricuspid Valve Repair Techniques: An Ex Vivo Study. Am J Physiol Heart Circ Physiol 2019;317:H1312-27. doi: 10.1152/ajpheart.00150.2019
[8] Salurso E, Jaworek M, Perico F, Frigelli M, RomagnoniC, Contino M, et al. Morphometric Characterization of an Ex Vivo Porcine Model of Functional Tricuspid Regurgitation. Ann Biomed Eng 2022;51:715-25. doi: 10.1007/s10439-022-03080-2
[9] Pasrija C, Quinn R, Ghoreishi M, Eperjesi T, Lai E, Gorman RC, et al. A Novel Quantitative Ex Vivo Model of Functional Mitral Regurgitation. Innovations (Phila) 2020;15:329-37. doi: 10.1177/1556984520930336
[10] Crick SJ, Sheppard MN, Ho SY, Gebstein L, AndersonRH. Anatomy of the pig Heart: Comparisons with Normal Human Cardiac Structure. J Anat 1998;193:105-19. doi: 10.1046/j.1469-7580.1998.19310105.x
[11] Karamali F, Hosseini S, Shojaeifard M, Mohammadi K, Kaviani R, Rezaei Y, et al. Tricuspid Valve Geometry in Patients with Functional Tricuspid Regurgitation: A Three-Dimensional Echocardiographic Study. Echocardiography 2020;37:867-75. doi: 10.1111/echo.14747
[12] Florescu DR, Muraru D, Florescu C, Volpato V, CaravitaS, Perger E, et al. Right Heart Chambers Geometry and Function in Patients with the Atrial and the Ventricular Phenotypes of Functional Tricuspid Regurgitation. Eur Heart J Cardiovasc Imaging 2022;23:930-40. doi: 10.1093/ehjci/jeab211
[13] Ton-Nu TT, Levine RA, Handschumacher MD, Dorer DJ, Yosefy C, Fan D, et al. Geometric determinants of Functional Tricuspid Regurgitation: Insights from 3-Dimensional Echocardiography. Circulation 2006;114:143-9. doi: 10.1161/CIRCULATIONAHA.106.611889
[14] Badano LP, Hahn R, Rodríguez-Zanella H, Araiza Garaygordobil D, Ochoa-Jimenez RC, Muraru D. Morphological Assessment of the Tricuspid Apparatus and Grading Regurgitation Severity in Patients with Functional Tricuspid Regurgitation: Thinking Outside the Box. JACC Cardiovasc Imaging 2019;12:652-64. doi: 10.1016/j.jcmg.2018.09.029
[15] Badano LP, Agricola E, Perez de Isla L, Gianfagna P, Zamorano JL. Evaluation of the Tricuspid Valve Morphology and Function by Transthoracic Real-Time Three-Dimensional Echocardiography. Eur J Echocardiogr 2009;10:477-84. doi: 10.1093/ejechocard/jep044
[16] Fukuda S, Saracino G, Matsumura Y, Daimon M, Tran H, Greenberg NL, et al. Three-Dimensional Geometry of the Tricuspid Annulus in Healthy Subjects and in Patients with Functional Tricuspid Regurgitation: A Real-Time, 3-Dimensional Echocardiographic Study. Circulation 2006;114 1 Suppl: I492-8. doi: 10.1161/CIRCULATIONAHA.105.000257
[17] Fukuda S, Gillinov AM, Song JM, Daimon M, Kongsaerepong V, Thomas JD, et al. Echocardiographic Insights into Atrial and Ventricular Mechanisms of Functional Tricuspid Regurgitation. Am Heart J 2006;152:1208-4. doi: 10.1016/j.ahj.2006.07.027
[18] Fukuda S, Gillinov AM, McCarthy PM, Stewart WJ, Song JM, Kihara T, et al. Determinants of Recurrent or Residual Functional Tricuspid Regurgitation after Tricuspid Annuloplasty. Circulation 2006;114 1 Suppl: I582-7 doi: 10.1161/CIRCULATIONAHA.105.001305
[19] Kabasawa M, Kohno H, Ishizaka T, Ishida K, Funabashi N, Kataoka A, et al. Assessment of Functional Tricuspid Regurgitation using 320-Detector-Row Multislice Computed Tomography: Risk Factor Analysis for Recurrent Regurgitation after Tricuspid Annuloplasty. J Thorac Cardiovasc Surg 2014;147:312-20. doi: 10.1016/j.jtcvs.2012.11.017
[20] Maslow A, Abisse S, Parikh L, Apruzzese P, Cilia L, Gleason P, et al. Echocardiographic Predictors of Tricuspid Ring Annuloplasty Repair Failure for Functional Tricuspid Regurgitation. J Cardiothorac Vasc Anesth 2019;33:2624-33. doi: 10.1053/j.jvca.2019.05.043
[21] Waziri F, Lyager Nielsen S, Michael Hasenkam J. Porcine Tricuspid Valve Anatomy and Human Compatibility: Relevance for Preclinical Validation of Novel Valve Interventions. J Heart Valve Dis 2016;25:596-605.
[22] Fawzy H, Fukamachi K, Mazer CD, Harrington A, Latter D, Bonneau D, et al. Complete Mapping of the Tricuspid Valve Apparatus using Three-Dimensional Sonomicrometry. J Thorac Cardiovasc Surg 2011;141:1037-43. doi: 10.1016/j.jtcvs.2010.05.039
[23] Maisano F, Taramasso M, Guidotti A, Redaelli A, Fiore GB, Baroni G, et al. Simulation of Functional Tricuspid Regurgitation using an Isolated Porcine Heart Model. J Heart Valve Dis 2011;20:657-63.
[24] Madhurapantula RS, Krell G, Morfin B, Roy R, Lister K, Orgel JP. Advanced Methodology and Preliminary Measurements of Molecular and Mechanical Properties of Heart Valves under Dynamic Strain. Int J Mol Sci 2020;21:763. doi: 10.3390/ijms21030763
[25] Roy R, Warren E Jr., Xu Y, Yow C, Madhurapantula RS, Orgel JP, et al. Functional Grading of a Transversely Isotropic Hyperelastic Model with Applications in Modeling Tricuspid and Mitral Valve Transition Regions. Int J Mol Sci 2020;21:6503. doi: 10.3390/ijms21186503